Identification of a First Order Model of Implanted Electrodes on the First SUAW Patient - LIRMM - Laboratoire d’Informatique, de Robotique et de Microélectronique de Montpellier
Communication Dans Un Congrès Année : 2002

Identification of a First Order Model of Implanted Electrodes on the First SUAW Patient

Résumé

Impedance evaluation of implanted electrodes is necessary to ensure that the implant can deliver the specified current, and to survey any failure due to the electrodes. To perform the identification in the implanted hardware, a simple, but sufficiently descriptive model, must be used because of the poor excitation input signal and in order to avoid complex identification algorithms. Data recorded on the first implanted SUAW patient give information on the real impedance of both neural and epimysial electrodes and an opportunity to validate simple first order models.
Fichier principal
Vignette du fichier
Identification_of_a_First_Order_Model_of_Implanted.pdf (271.54 Ko) Télécharger le fichier
Origine Fichiers produits par l'(les) auteur(s)
Loading...

Dates et versions

lirmm-00268646 , version 1 (04-10-2019)

Identifiants

  • HAL Id : lirmm-00268646 , version 1

Citer

David Guiraud, Jean-Louis Divoux, Pierre Rabischong. Identification of a First Order Model of Implanted Electrodes on the First SUAW Patient. IFESS: International Functional Electrical Stimulation Society, Jun 2002, Ljubljana, Slovenia. pp.214-216. ⟨lirmm-00268646⟩