Conference Papers Year : 2011

An Electrical Test Method for MEMS Convective Accelerometers: Development and Evaluation

Abstract

In this paper, an alternative test method for MEMS convective accelerometers is presented. It is first demonstrated that device sensitivity can be determined without the use of physical test stimuli by simple electrical measurements. Using a previously developed behavioral model that allows efficient Monte-Carlo simulations, we have established a good correlation between electrical test parameters and device sensitivity. Proposed test method is finally evaluated for different strategies that privilege yield, fault coverage or test efficiency.
Fichier principal
Vignette du fichier
05763137.pdf (690.68 Ko) Télécharger le fichier
Origin Publisher files allowed on an open archive
Loading...

Dates and versions

lirmm-00702754 , version 1 (06-10-2019)

Identifiers

Cite

Ahmed Rekik, Florence Azaïs, Norbert Dumas, Frédérick Mailly, Pascal Nouet. An Electrical Test Method for MEMS Convective Accelerometers: Development and Evaluation. DATE: Design, Automation and Test in Europe, Mar 2011, Grenoble, France. ⟨10.1109/DATE.2011.5763137⟩. ⟨lirmm-00702754⟩