Efficient calibration of contact-less resonant bio-sensor affected by operating conditions

Anthony Deluthault 1 Vincent Kerzérho 1 Serge Bernard 1 Fabien Soulier 1 Philippe Cauvet 2
1 SysMIC - Conception et Test de Systèmes MICroélectroniques
LIRMM - Laboratoire d'Informatique de Robotique et de Microélectronique de Montpellier
Abstract : In this paper, a new model-based method is presented for the development of calibration technique of inductively coupled biosensors. This method is applied to the case of a lens-based intra-ocular pressure sensor. A new performance indicator is proposed. It is the 2 nd-order coefficient of the single-variable quadratic polynomial regression approximating the Bode phase plot of the system. This new indicator is reliable to significant distance and angle variations between primary and secondary and to small signal-to-noise-ratio.
Complete list of metadatas

Cited literature [9 references]  Display  Hide  Download

https://hal-lirmm.ccsd.cnrs.fr/lirmm-01430618
Contributor : Caroline Lebrun <>
Submitted on : Tuesday, January 10, 2017 - 10:09:46 AM
Last modification on : Tuesday, April 2, 2019 - 10:44:01 AM
Long-term archiving on : Tuesday, April 11, 2017 - 1:54:18 PM

File

07524219.pdf
Publisher files allowed on an open archive

Identifiers

Citation

Anthony Deluthault, Vincent Kerzérho, Serge Bernard, Fabien Soulier, Philippe Cauvet. Efficient calibration of contact-less resonant bio-sensor affected by operating conditions. IMSTW: International Mixed-Signal Testing Workshop, Jul 2016, Sant Feliu de Guixols, Spain. pp.1-6, ⟨10.1109/IMS3TW.2016.7524219⟩. ⟨lirmm-01430618⟩