Continuous Intra Ocular Pressure Measurement Sensor for Glaucoma Diagnostic

Abstract : Glaucoma is an ocular pathology usually associated with an increase in Intra Ocular Pressure (IOP). In this study, we are developing disposable eye lenses including a specific micro fabricated pressure sensor to measure IOP all day long. The information data will be wireless transferred via magnetic coupling to an external receiver. Our first work deals with the sensor design and fabrication. Simulation results based on classical electronic circuit tools will be presented and lead to several sensor solutions working at different RF frequencies. The fabrication process of the first sensors will also be described and an early IOP characterisation set-up will be presented to try to later quantify the sensor sensitivity, with IOP variations measured in the range 20 to 70mmHg.
Type de document :
Communication dans un congrès
6th World Congress of Biomechanics, Aug 2010, Singapore. 31, pp.1282-1285, 2010
Liste complète des métadonnées

https://hal-lirmm.ccsd.cnrs.fr/lirmm-00808458
Contributeur : Serge Bernard <>
Soumis le : vendredi 5 avril 2013 - 14:36:04
Dernière modification le : jeudi 13 décembre 2018 - 01:48:42

Identifiants

  • HAL Id : lirmm-00808458, version 1

Citation

Gaelle Lissorgues, Lionel Rousseau, Patrick Poulichet, Laurie Valbin, Serge Picaud, et al.. Continuous Intra Ocular Pressure Measurement Sensor for Glaucoma Diagnostic. 6th World Congress of Biomechanics, Aug 2010, Singapore. 31, pp.1282-1285, 2010. 〈lirmm-00808458〉

Partager

Métriques

Consultations de la notice

320