New Calibration Technique of Contact-less Resonant Biosensor - LIRMM - Laboratoire d’Informatique, de Robotique et de Microélectronique de Montpellier Accéder directement au contenu
Article Dans Une Revue Journal of Electronic Testing: : Theory and Applications Année : 2017

New Calibration Technique of Contact-less Resonant Biosensor

Anthony Deluthault
  • Fonction : Auteur
  • PersonId : 961591
Vincent Kerzérho
Serge Bernard
Fabien Soulier
Philippe Cauvet
  • Fonction : Auteur
  • PersonId : 866301

Résumé

This paper addresses the problem of calibrating magnetically-coupled passive sensors for medical applications. We developed an analytical model of such sensor including the impact of some disturbances (distance and angle between antennas). This model is valid for sensors with different dimensions of antennas. Thanks to this model, the ideal indicator of the disturbances for calibration can be identified. It has been applied to the case study of Intra-Ocular Pressure (IOP) sensor. A new indicator (the slope of the baseline of the Bode phase plot) has been chosen.This new indicator is reliable to significant distance and angle variations between primary and secondary.
Fichier non déposé

Dates et versions

lirmm-01925598 , version 1 (16-11-2018)

Identifiants

Citer

Anthony Deluthault, Vincent Kerzérho, Serge Bernard, Fabien Soulier, Philippe Cauvet. New Calibration Technique of Contact-less Resonant Biosensor. Journal of Electronic Testing: : Theory and Applications, 2017, 33 (3), pp.365-375. ⟨10.1007/s10836-017-5660-2⟩. ⟨lirmm-01925598⟩
86 Consultations
0 Téléchargements

Altmetric

Partager

Gmail Facebook X LinkedIn More